
APPLIED OPERATIONS
/Protocol processor
IV.
OPERATION INSTRUCTIONS
IV .
2 - 1 5 7
Page
mask which is displayed on FLUOVIEW software or to disable the mask by
selecting “None”.
[REX Mask] appears only when AOTF (FV5-COMBA) is equipped.
PMT Voltage: Enter the PMT voltage of each channel. The setting is 1 to 1100, and the
unit is V.
Gain: Enter the gain of each channel. The setting is 1.0 to 10.0.
Offset: Enter the offset of each channel. The setting is 0 to 100, the unit is %.
XY Stage Track No: When multi-point time lapse software is used, track number of
multi-point time lapse software that meets to this track will be
displayed. When multi-point time lapse software is not used, the
display does not appear.
Zstatus: Enter Z stage information.
Start Pos: Enter the Z-scanning start position of cross section of which
observation is required.
Stop Pos: Enter the Z-scanning stop position of cross section of which
observation is required.
Step Size: Enter the number of steps in Z direction.
Slices: Enter the number of image slices to be acquired Save FileName: Input file name
for acquired image. Image would be saved under path of [Image
File Name:].
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......