APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 2 0
Page
When the light incident to photomultiplier tube is too bright or PMT
voltage is set to a high voltage, PMT Over warning may be displayed
as shown below.
This warning is displayed to protect photomultiplier tube when the
light incident to it exceeds a certain level. When it is displayed,
decrease PMT voltage setting.
[Hi-Lo] LUT can be utilized to adjust image intensity easily.
1. Click <LUT> button from tool bar located at bottom left of screen.
Dialog box – [Color Tool] will appear.
2. Click [Hi-Lo] LUT from group box of [Standard Color LUTs].
3. The intensity value 0 is colored with Blue, and the maximum intensity
is colored with Red.
4. When there is noises in the image, different colors may appear in its
background. When the intensity is saturated, the portion will be colored
with Red. Based on this coloring, you may adjust the image intensity.
[Hi-Lo] LUT switching can be done by use of Hot key even if during image
acquisition.
The image channel can be assigned in [Acquire] panel
The LUT of the assigned channel can be switched to [Hi-Lo] LUT by
pressing
㪟
+
㪚㫋㫉㫃
. By pressing
㪟
+
㪚㫋㫉㫃
again, LUT
returns to the previous. The LUT of all channels can be switched to [Hi-Lo]
LUT by pressing
㪟
+
㪪㪿㫀㪽㫋
㪚㫋㫉㫃
+
.
NOTE
<LUT> button
[Hi-Lo]LUT
TIP
TIP
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......