Engage the filter by setting the [BARRIER FILTERS] lever of the scan unit. (Setting
the lever to “IN” engages the filter.)
Also see section 2-2-1-2, “Configuring the Scan Unit” and follow the instructions
given by the [Microscope Configuration] window.
z
Turn off the transmitted light bulb of the microscope.
1. From the page tabs on the bottom right of the [Acquire] panel, select the
[Settings] sub-panel.
Fig. 2-21 [Settings] Sub-panel
2. Click the <Trans. Lamp> button to turn the lamp off (set the button to the non-
pressed status).
<Trans. Lamp> button
Turns off the transmitted light bulb of the
microscope.
(The lamp lights when the button is in
the pressed status.)