Appendix I List of Functions in the [Active Overlays] Dialog Box
/Intensity Data
IV.
OPERATION INSTRUCTIONS
IV .
I - 4
Page
Appendix I-2 Intensity Data
The intensity value can be displayed.
When images are overlapped, the intensity value of each image is accompanied with the
channel number placed after it.
«Syntax»
<intensity[ [hotspot] [raw/calibrated] value [units]]>
Arguments inside [ ] can be omitted.
If [raw/calibrated] is omitted, the same setting as when [calibrated] is specified will be
applied.
«Setup Procedure»
1. Enter “intensity” as the first character inside < >.
2. Add the following characters to set the display method.
hotspot
: Display measurement points with “+” markings.
raw
: Display data in values between 0 and 4095.
calibrated
: Display data in numerical values (µm).
units
: Display the unit if this has been set.
(Same as the concentration computation in FV-TIEMPO)
«Examples»
<Input character strings>
<Displayed strings (** represent figures.)>
<intensity raw value> ..................................**
<intensity hotspot raw value> ......................***(1),**(2)
(When images are overlapped)
The detailed display procedure is described in section 2-11-2, “Viewing the
Intensity Value of Image” in Volume [OPERATION].
TIP
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......