Product Specifications
A-16
A.5.6 IBP
Standards
Meet standard of EN60601-2-34/IEC60601-2-34.
Technique Direct
invasive
measurement
IBP
ART
0 to 300 mmHg
PA
-6 to 120 mmHg
CVP/RAP/LAP/ICP
-10 to 40 mmHg
Measurement range
P1/P2
-50 to 300 mmHg
Resolution 1
mmHg
Accuracy
±2% or ±1mmHg, whichever is greater
Refreshing rate
1s
Pressure transducer
Excitement voltage
5 VDC, ±2%
Sensitivity 5
uV/V/mmHg
Impedance range
300 to 3000
Ω
A.5.7 C.O.
Measurement method
Thermodilution method
C.O.:
0.1 to 20 L/min
BT:
23 to 43
°
C
Measurement range
TI:
0 to 27
°
C
Resolution
C.O.:
BT, TI:
0.1 L/min
0.1
°
C
Accuracy
C.O.:
BT, TI:
±5% or ±0.1 L /min
±0.1
°
C (without sensor)
Alarm range
BT:
23 to 43
°
C