Product Specifications
A-19
A.5.9 AG
Standards
Meet standard of ISO 21647
Technique Infrared
absorption
Warm-up time
Warming up status:
Ready-to-measure status:
45 s
10 min
Sample flow rate
Adult, pediatric:
Neonate:
Accuracy:
120, 150, 200 ml/min
70, 90, 120 ml/min
±10 ml/min or ±10%, whichever is greater
Measurement range
CO
2
:
O
2
:
N
2
O:
Des:
Sev:
Enf:
Iso:
Hal:
awRR:
0 to 30%
0 to 100%
0 to 100%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
2 to 100 rpm
Resolution
CO
2
:
awRR:
1 mmHg
1 rpm
Accuracy
(warming-up status)
CO
2
:
N
2
O:
Other anesthetic gases:
±0.3%
ABS
±(8%
REF
+2%
ABS
)
8%
REF
Gases
Range (%
REL
)
Accuracy (%
ABS
)
0 to 1
±0.1
1 to 5
±0.2
5 to 7
±0.3
7 to 10
±0.5
CO
2
>10 Not
specified
0 to 20
±2
N
2
O
20 to 100
±3
0 to 25
±1
25 to 80
±2
O
2
80 to 100
±3
0 to 1
±0.15
1 to 5
±0.2
Accuracy
(ready-to-measure
status)
Des
5 to 10
±0.4
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...