Alarms
6-5
6.4 Alarm Tone Configuration
6.4.1 Changing the Alarm Volume
1. Select
the
[
Alarm Setup
] QuickKey, or [
Main Menu
]
→
[
Alarm Setup >>
].
2. Select
[
Alarm Volume
] and then select the appropriate volume. 1 is minimum volume,
and 10 the maximum volume.
6.4.2 Switching the Alarm Sound On/Off
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password.
2. Select
[
Alarm Setup >>
] to enter the [
Alarm Setup
]
menu.
3. Select
[
Alarm Sound
] and toggle between [
On
] and [
Off
].
6.4.3 Setting the Interval between Alarm Sounds
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password.
2. Select
[
Alarm Setup >>
] to enter the [
Alarm Setup
]
menu.
3. Select
[
High Alarm Interval (s)
], [
Med Alarm Interval (s)
] and [
Low Alarm Interval
(s)
] in turn and then select the appropriate settings.
WARNING
z
When the alarm sound is switched off, the patient monitor will give no audible
alarm tones even if a new alarm occurs. Therefore the user should be very careful
about whether to switch off the alarm sound or not.
z
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the patient.
Always keep the patient under close surveillance.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...