Monitoring Cardiac Output
12-5
6. When you see the message [
Ready for New Measurement
], select the [
Start
] button
and then inject the solution within 4 seconds. As shown in the figure above, during the
measurement, the currently measured thermodilution curve is displayed on the upper
part. At the end of the measurement, the thermodilution curve is transferred to one of
the 6 measurement windows on the lower part and the monitor prompts you to wait for
a certain period of time before starting a new measurement.
7. When you see the message [
Ready for New Measurement
], repeat step 5 until you
have completed the measurements you want to perform. A maximum of 6
measurements can be stored. If you perform more than six measurements without
rejecting any, the oldest will automatically be deleted when a seventh curve is stored.
Select from the 6 measurement curves and the system will automatically calculate and
display the averaged C.O. and C.I. values.
When injecting, the stopcock to the PA catheter is open and the stopcock to the injectate
solution is closed. After the measurement is completed, turn off the stopcock to the PA
catheter and turn on the stopcock to the injectate solution, and then draw the injectate
solution into the injectate syringe.
In the buttons area, you can:
Select [
Start
] to start a C.O. measurement.
Select
Stop
] to stop the current measurement.
Select [
Cancel
] during a measurement to cancel the measurement. Selecting it after a
measurement deletes the measured results.
Select [
X-Scale
] to adjust the scale of the X-axis. Options for scale range are 30 s and
60 s.
Select [
Y-Scale
] to adjust the scale of the Y-axis. Options for scale range are 0.5ºC,
1ºC and 2.0ºC.
Select [
Record
] to print out the most recently-measured curve and numerics by the
recorder.
Select [
C.O. Setup >>
] to access the [
C.O. Setup
] menu.
Select [
Calculations >>
] to access the [
Hemodynamic Calculation
] menu.
NOTE
z
During the cardiac output measurement, blood temperature alarms are inactive.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...