Monitoring ECG/Resp
7-13
7.6 About ST Monitoring
ST segment analysis is only for use on adult patients. The default setting is off.
ST segment analysis calculates ST segment elevations and depressions for individual
leads and then displays them as numerics in the ST1 and ST2 areas.
A positive value indicates ST segment elevation; a negative value indicates ST segment
depression.
Measurement unit of the ST segment: mV.
Measurement range of the ST segment: -2.0 mV to +2.0 mV.
7.6.1 Switching ST On and Off
To switch ST monitoring on or off:
1. In
the
[
ECG Setup
] menu, select [
ST Analysis >>
].
2. Select
[
ST Analysis
] to toggle between [
On
] and [
Off
].
When ST monitoring is switched on, the patient monitor selects the [
Diagnostic
] filter mode
automatically. You can also select the [
Monitor
] or [
Surgery
] filter mode. But in that case,
the ST numerics may be severely distorted.
By selecting the ST parameter window, you can enter the [
ST Analysis
] menu.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...