4-1
4
Managing Patients
4.1 Admitting a Patient
The patient monitor displays physiological data and stores them in the trends as soon as a
patient is connected. This allows you to monitor a patient that is not yet admitted. However,
it is recommended that you fully admit a patient so that you can clearly identify your patient,
on recordings, reports and networking devices.
To admit a patient:
1. Select
the
[
Patient Setup
] QuickKey, or [
Main Menu
]
→
[
Patient Setup >>
].
2. Select
[
Discharge Patient
] to clear any previous patient data. If you do not erase data
from the previous patient, the new patient’s data will be saved into the data of the
previous patient. The monitor makes no distinction between the old and the new
patient data.
3. If
[
Discharge Patient
] button appears dimmed, directly select [
Admit Patient
] and
then select:
[
Yes
] to apply the data saved in the patient monitor to the new patient, or
[
No
]
to clear the data saved in the patient monitor.
4. In
the
[
Patient Demographics
] menu, enter the demographic details, of which:
[
Patient Cat.
] determines the way your patient monitor processes and calculates
some measurements, and what safety and alarm limits are applied for your patient.
[
Paced
] determines whether to show pace pulse marks on the ECG waveform.
When the
[
Paced
]
is set to [
No
], pace pulse marks are not shown in the ECG
waveform.
5. Select
[
Ok
].
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...