Product Specifications
A-13
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate
1s
Low Perfusion
Conditions
Pulse amplitude: >0.02%
Light penetration: >5%
SpO
2
accuracy
±2%
PR accuracy
±3 bpm
Nellcor SpO
2
Module
Parameter
Specifications
Sensor
Range
Precision*
MAX-A, MAX-AL, MAX-N
MAX-P, MAX-I, MAX-FAST
70 to 100%
0% to 69%
±2%
Not specified.
OxiCliq A, OxiCliq N
OxiCliq P, OxiCliq I
70 to 100%
0% to 69%
±2.5%
Not specified.
D-YS, DS-100A, OXI-A/N,
OXI-P/I
70 to 100%
0% to 69%
±3%
Not specified.
SpO
2
measurement range and
accuracy
MAX-R, D-YSE, D-YSPD
70 to 100%
0% to 69%
±3.5%
Not specified.
20 to 250 bpm:
±3 bpm
PR measurement range and
accuracy
251 to 300 bpm:
Not specified.
Refreshing rate
1s
*: When the SpO
2
sensor is applied for neonatal patients as indicated, the specified accuracy range
is increased by ±1%, to compensate for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...