Factory Defaults
C-9
C.12 AG Setup
AG Setup
Factory default settings
Alarm On
Alm Lev
Med
Alm Rec
Off
Apnea Time
20 s
Flow Rate
Low
O2 Compen
Off
Operating Mode
Standby
Auto Standby
Off
Wave Type (CO2)
Draw
Sweep 12.5
mm/s
Alarm Limits
Setup
Adult/pediatric/neonate
Alarm Limits
Setup
Adult/pediatric/neonate
awRR High
30/30/100
awRR Low
8/8/30
EtCO2 High
50/50/45
EtO2 High
684
EtCO2 Low
15/15/30
EtO2 Low
137
FiCO2 High
4
FiO2 High
669
FiCO2 Low
0
FiO2 Low
137
EtN2O High
55
EtHal High
3.0
EtN2O Low
0
EtHal Low
0.0
FiN2O High
53
FiHal High
2.0
FiN2O Low
0
FiHal Low
0.0
EtEnf High
3.0
EtIso High
3.0
EtEnf Low
0.0
EtIso Low
0.0
EtEnf High
2.0
EtIso High
2.0
EtEnf Low
0.0
EtIso Low
0.0
EtSev High
6.0
EtDes High
8.0
EtSev Low
0.0
EtDes Low
0.0
FiSev High
5.0
FiDes High
6.0
FiSev Low
0.0
FiDes Low
0.0
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...