Product Specifications
A-7
A.3.6 Monitor Interface Specifications
Power
1 AC power input connector
2
Wire network
RJ45 connector, 100-Base-TX, IEEE 802.3
USB
10 connectors, USB 1.1
SMR connector
1 connector, not standard USB
CF
50-pin CF revision 2.0 connector
Video interface
1 connector, standard DVI-D
Auxiliary output
1 connector, standard BNC, the common connector of nurse call
signals and analog output signals
Equipotential Grounding
Terminal
1
Defib Sync Connector
1 connector, RJ11
A.3.7 Outputs
Auxiliary Output
Standard
Meets the requirements of EC60601-1 for short-circuit
protection and leakage current
Output impedance
50
Ω
rating
ECG Analog Output
Bandwidth
(-3dB; reference frequency:
10Hz)
Diagnostic mode:
Monitor mode:
Surgical mode:
0.05 to 100 Hz
0.5 to 40 Hz
1 to 20 Hz
Output signal range
-2.5 to 2.5 V
Max transmission delay
25 ms (in diagnostic mode, and with Notch off)
Sensitivity
1V/mV ±5%
PACE rejection/enhancement No pace rejection or enhancement
IBP Analog Output
Bandwidth (-3dB; reference
frequency:1Hz)
0 to 12.5 Hz
Max transmission delay
55 ms (with Notch off)
Sensitivity
1 V/100 mmHg ±5%
Nurse Call Signal
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...