Monitoring Carbon Dioxide
13-4
prevents them from entering the module. If the collected water reaches to a certain
amount, you should drain it to avoid blocking the airway.
z
The watertrap has a filter preventing bacterium, water and secretions from
entering the module. After a long-term use, dust or other substances may
compromise the performance of the filter or even block the airway. In this case,
replace the watertrap. Replacing the watertrap once a month is recommended.
13.3.2 Using a Microstream CO
2
Module
1. Connect the sampling line to the module and then connect the CO
2
components as
shown below.
2. By default, the microstream CO
2
module is in measure mode. The message [
CO2
Sensor Warmup
] appears on the screen when the CO
2
module is plugged.
3. After warm-up, you can perform CO
2
measurements.
Sampling line
Connector for
sampling line
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...