Monitoring Carbon Dioxide
13-10
13.7 Zeroing the Sensor
The zero calibration eliminates the effect of baseline drift during CO
2
measurement exerted
on the readings and therefore maintains the accuracy of the CO
2
measurements.
13.7.1 For Sidestream and Microstream CO
2
Modules
For sidestream and microstream CO
2
modules, a zero calibration is carried out automatically
when necessary. You can also start a manual zero calibration if necessary. To manually start a
zero calibration, select [
Start Zero Cal.
] in the [
CO2 Setup
] menu. Disconnecting the
patient airway is not required when performing a zero calibration.
13.7.2 For Mainstream CO
2
Modules
For mainstream CO
2
modules, zero the sensor whenever:
A new adapter is used;
You reconnect the sensor to the module;
You see the message [
CO2 Zero Required
]. In this case, check the airway adapter for
any blockage, e.g. mucus, etc. If a blockage is detected, clear or replace the adapter.
To zero the sensor, follow this procedure:
1. Connect the sensor to the module.
2. In
the
[
CO2 Setup
] menu, set the [
Operating Mode
] to [
Measure
]. The message [
CO2
Sensor Warmup
] is displayed.
3. After warm-up is finished, connect the sensor to a clean, dry airway adapter. The adapter
should be vent to the air and isolated from CO
2
sources, such as ventilator, the patient’s
breathing, your own breathing, etc.
4. Select
[
Start Zero Cal.
] in the [
CO2 Setup
] menu. The message [
CO2 Zero Running
]
is displayed.
5. It takes about 15 to 20 seconds. The message disappears when the zero calibration is
completed.
WARNING
z
When perform a zero calibration during the measurement, disconnect the
transducer from the patient’s airway first.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...