Monitoring NIBP
9-7
You can not display multiple sets of NIBP measurements in some screens such as the big
numerics screen and the 12-lead ECG analysis screen.
9.7.3 Setting the Pressure Unit
In the [
NIBP Setup
] menu, select [
Unit
] and toggle between [
mmHg
] and [
kPa
].
9.8 Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel
and therefore help venous puncture.
1. In
the
[
NIBP Setup
] menu, verify that the [
Cuff Press.
] value is appropriate. Change it
if necessary.
2. Select
[
VeniPuncture
].
3. Puncture vein and draw blood sample.
4. Reselect either [
VeniPuncture
], or the
hardkey on the monitor’s front, or the
on-screen QuickKey to deflate the cuff. The cuff deflates automatically after a set time if
you do not deflate it.
During measurement, the NIBP display shows the inflation pressure of the cuff and the
remaining time in venous puncture mode.
9.9 Resetting NIBP
If the blood pressure pump works incorrectly but the monitor does not alarm for it, you can
check the pump by resetting it. To reset the pump, select [
Reset
] from [
NIBP Setup
].
9.10 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required once
every two years or when you doubt the measured NIBP. If the test failed, corresponding
prompt messages will be given. If no message is displayed, it means no leakage is detected.
Tools required:
An adult cuff
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...