Monitoring ICG
15-6
15.7 Changing ICG Settings
15.7.1 ICG Averaging
The ICG value is the average of multiple measurements. You can select an interval (heart
beats) for averaging ICG.
1. Select the ICG parameter window to enter the [
ICG Setup
] menu.
2. Select
[
Averaging
] and then select the appropriate setting. The greater the averaging
interval is, the less the ICG value is affected by human interference. Vice versa.
15.7.2 Selecting Secondary Parameters
C.O., SVR and TFC are the default three secondary parameters. You can also select your
desired secondary parameter for display.
1. Select the ICG parameter window to enter the [
ICG Setup
] menu.
2. Select
[
Change Secondary Parameters >>
].
3. Select
three
parameters from the popup menu.
15.7.3 Checking Sensors
During ICG monitoring, the ICG sensors should be checked regularly to ensure that no sensor
becomes disconnected. During sensor checking, the ICG waveform is displayed as a straight
line and the message [
ICG Sensor Check
] is displayed. Once a disconnected sensor is
detected, a prompt message in which the sensor’s application site is indicated will be
displayed.
To initiate a sensor check:
In the [
ICG Setup
] menu, select [
Check Sensor
].
Press the
hardkey.
15.7.4 Changing the ICG Wave Speed
1. Select the ICG wave to enter the [
ICG Waveform
] menu.
2. Select
[
Sweep
] and then select the appropriate setting. The faster the wave sweeps, the
wider the wave is.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...