Monitoring IBP
11-3
WARNING
z
If measuring intracranial pressure (ICP) with a sitting patient, level the transducer
with the top of the patient’s ear. Incorrect leveling may give incorrect values.
11.4 Understanding the IBP Display
The IBP measurement is displayed on the monitor as a waveform and numeric pressures. The
figure below shows the waveform and numerics for the Art pressure. For different pressures,
this display may be slightly different.
1.
Waveform
2. Systolic
pressure
3. Diastolic
pressure
4. Mean
pressure
5. Pressure
unit
For some pressures, the parameter window may show the mean pressure only. For different
pressures, their default unit may be different. If the Art and ICP pressures are measured
simultaneously, the parameter window of the ICP display will show numeric CPP, which is
obtained by subtracting ICP from the Art mean.
1
2
3
4
5
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...