Product Specifications
A-20
10 to 15
±0.6
15 to 18
±1
>18 Not
specified
0 to 1
±0.15
1 to 5
±0.2
5 to 8
±0.4
Sev
>8 Not
specified
0 to 1
±0.15
1 to 5
±0.2
Enf, Iso, Hal
>5 Not
specified
2 to 60 rpm
±1 rpm
awRR
>60 rpm
Not specified
Apnea alarm time
20 to 40 s
Refreshing rate
1s
CO
2
≤
250ms (fall time: 200ms)
N
2
O
≤
250ms
O
2
≤
600ms
Hal, Iso, Sev, Des
≤
300ms
Rise time*
Enf
≤
350ms
CO
2
≤
250ms (fall time: 200ms)
N
2
O
≤
250ms
O
2
≤
500ms
Hal, Iso, Sev, Des
≤
300ms
Rise time**
Enf
≤
350ms
Delay time
<4s
*: 10 % to 90 %, gas sample flow rate 120ml/min, using the DRYLINE™ watertrap and neonatal
DRYLINE™ sampling line (2.5m)
**: 10 % to 90 %, gas sample flow rate 200ml/min, using the DRYLINE™ water trap and adult
DRYLINE™ sampling line (2.5m).
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...