Product Specifications
A-9
A.5 Measurement Specifications
A.5.1 ECG
ECG
Standards
Meet standards of EC11, EC13, EN60601-2-27/IEC60601-2-27
and IEC60601-2-25
Lead set
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6
ECG standard
AHA, IEC
Display sensitivity
1.25 mm/mV (X0.125), 2.5 mm/mV (X0.25), 5 mm/mV (X0.5),
10 mm/mV (X1), 20 mm/mV (X2), Auto
Sweep speed
12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth (-3dB)
Diagnostic mode:
Monitor mode:
Surgical mode:
0.05 to 150 Hz
0.5 to 40 Hz
1 to 20 Hz
Common mode rejection ratio
(with Notch off)
Diagnostic mode:
Monitor mode:
Surgical mode:
≥
90 dB
≥
105 dB
≥
105 dB
Notch
50/60 HZ
Monitor and surgical mode: Notch turns on automatically.
Diagnostic mode: Notch is turned on/off manually
Differential input impedance
≥
5 M
Ω
Input signal range
±8 mV (peak-to-peak value)
Electrode offset potential
tolerance
±500 mV
Baseline recovery time
<5s (after defibrillation)
Patient leakage current
<10 uA
Calibration signal
1mV (peak-to-peak value)
HR
3-, 5-, and 12-lead
ECG
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Measurement range
12-lead analysis
Adult, pediatric:
30 to 120 bpm
Resolution 1
bpm
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...