Product Specifications
A-22
A.5.12 RM
Technique
Flow sensor
Frequency response
≥
30Hz
Dead space
<11ml
Flow
Measurement range
Adult/pediatric:
Neonate:
± (2 to 120) L/min
± (0.5 to 30) L/min
Accuracy
25 ml/s or ±10% of the reading, whichever is greater
Resolution
0.1 L/min (minimum)
Paw
Measurement range
-20 to 120 cmH
2
O
Accuracy
±3% of the reading
Resolution 0.1
cmH
2
O
MVe/MVi
Measurement range
Adult, Pediatric:
Neonatal:
2 to 60 L/min
0.5 to 15 L/min
Accuracy ±5%×reading
TVe/TVi
Measurement range
Adult, Pediatric:
Neonatal:
100 to 1500 ml
20 to 500 ml
Resolution 1
ml
Accuracy
±15 ml or ±10%×reading, whichever is greater.
RR (RM)
Measurement range
4 to 120 rpm
Accuracy
4 to 99 rpm
100 to 120 rpm
±1 rpm
±2 rpm
Calculated Parameters
Measurement range
Measurement accuracy
I:E
12:1 to 1:12
FEV1.0%
0 to 100%
Pmean
0 to 120 cmH
2
O
TV
20 to 1500 ml
MV
2 to 60 L
PEEP
0 to 120 cmH
2
O
<±5% of the reading
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...