Alarm Messages
D-3
D.2 Technical Alarm Messages
Measure
ment
Alarm message
A
B
Cause and solution
XX SelfTest Err
√
×
XX Init Err
√
×
XX Init Err N
√
×
N is within 1 to 8
XX Comm Err
√
×
XX Comm Stop
×
×
An error occurred to the XX module, or there is
a problem with the communications between the
module and the monitor. Re-plug the module
and restart the monitor, or plug the module into
another monitor.
XX Limit Err
×
×
XX parameter limit is accidentally changed.
Contact your service personnel.
XX
XX Overrange
×
×
The measured XX value is not within the
specified range for XX measurement. Contact
your service personnel.
MPM 12V Err
×
×
MPM
MPM 5V Err
×
×
An error occurred to the power supply part of
the MPM module. Contact your service
personnel.
ECG Lead Off
×
√
ECG YY Lead Off ECG
×
√
Note: YY represents the leadwires,
V (V1, V2, V3, V4, V5, V6,), LL,
LA, RA, as per AHA standard, or C
(C1, C2, C3, C4, C5, C6), F, L and R
as per IEC standard.
The electrode has become detached from the
patient or the lead wire has become
disconnected from the adapter cable. Check the
connection of the electrodes and leadwires.
ECG Noisy
√
×
The ECG signal is noisy. Check for any possible
sources of signal noise form the area around the
cable and electrode, and check the patient for
great motion.
ECG
ECG Signal Invalid
×
√
The ECG signal is saturated. Check the
leadwires.
Resp Resp
Disturbed
√
×
The respiration circuit is disturbed. Restart the
monitor.
Temp Cal. Err
×
×
A calibration failed. Restart the monitor.
T1 Sensor Off
√
×
Temp
T2 Sensor Off
√
×
The Temp sensor has become detached from the
patient or the module. Check the sensor
connections.
SpO
2
SpO2
Sensor
Off
√
×
The SpO
2
sensor has become detached from the
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...