Monitoring Carbon Dioxide
13-5
13.3.3 Using a Mainstream CO
2
Module
1. Connect the sensor to the module.
2. By default, the mainstream CO
2
module is in standby mode. The message [
CO2
Standby
] appears on the screen when the CO
2
module is plugged. To set CO
2
module to
measure mode, select the
hardkey on the module to switch to the [
Measure
]
mode, or select the CO
2
parameter window
→
[
Operating Mode
]
→
[
Measure
]. The
message [
CO2 Sensor Warmup
] is displayed.
3. After warm-up is finished, connect the transducer to the airway adapter.
4. Perform a zero calibration per the
Zeroing the Sensor
section.
5. After the zero calibration is finished, connect the airway as shown below.
5. Make sure there are no leakages in the airway and then start a measurement.
NOTE
z
Always position the sensor with the adapter in an upright position to avoid
collection of fluids on the windows of the adapter. Large concentrations of fluids at
this point will obstruct gas analysis.
Airway adapter
Sensor
Connect to patient
Connect to monitor
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...