Monitoring Carbon Dioxide
13-7
13.4.4 Setting up CO
2
Compensations
WARNING
z
Make sure that the appropriate compensations are used. Inappropriate
compensations may cause inaccurate measurement values and result in
misdiagnosis.
For the sidestream CO
2
module, in the [
CO2 Setup
] menu, respectively select:
[
O2 Compen
] and toggle between [
On
] and [
Off
]. When the amount of O
2
in the
ventilation gas mixture is greater than 50%, you’d better switch this compensation on.
[
N2O Compen
] and toggle between [
On
] and [
Off
]. When the amount of N
2
O in the
ventilation gas mixture is greater than 30%, you’d better switch this compensation on.
[
Des Compen
] and toggle between [
On
] and [
Off
]. If desflurane is present in the
ventilation gas mixture, you’d better switch this compensation on.
For the microstream CO
2
module, gas compensations are not required.
For the mainstream CO
2
module, in the [
CO2 Setup
] menu, respectively select:
[
Balance Gas
] and toggle between [
Room Air
] and [
N2O
]. Select [
Room Air
] when
air predominates in the ventilation gas mixture and select [
N2O
] when N
2
O
predominates in the ventilation gas mixture.
[
O2 Compen
] and then select [
Off
] or an appropriate setting according to the amount of
O
2
in the ventilation gas mixture. When the amount of O
2
is less than 30%, you’d better
switch this compensation off.
[
AG Compen
] and enter the concentration of anesthetic gas present in the ventilation
gas mixture. This could compensate for the effect of AG on the readings.
13.4.5 Humidity Compensation for Side/Microsctream CO
2
Readings
Sidestream and microstream CO
2
modules are configured to compensate CO
2
readings for
either Body Temperature and Pressure, Saturated Gas (BTPS), to account for humidity in the
patient’s breath, or Ambient Temperature and Pressure, Dry Gas (ATPD).
1. In
the
[
CO2 Setup
] menu, select [
Humidity Compen
].
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...