Monitoring ECG/Resp
7-22
In auto detection mode, the patient monitor adjusts the detection level automatically,
depending on the wave height and the presence of cardiac artifact. Note that in auto
detection mode, the detection level (a dotted line) is not displayed on the waveform.
Use auto detection mode for situations where:
The respiration rate is not close to the heart rate.
Breathing is spontaneous, with or without continuous positive airway pressure
(CPAP).
Patients are ventilated, except patients with intermittent mandatory ventilation
(IMV).
In manual detection mode, you adjust the dotted detection level line to the desired level
by selecting [
Upper Line
] or [
Lower Line
] and then selecting
or
beside them.
Once set, the detection level will not adapt automatically to different respiration depths.
It is important to remember that if the depth of breathing changes, you may need to
change the detection level.
Use manual detection mode for situations where:
The respiration rate and the heart rate are close.
Patients have intermittent mandatory ventilation.
Respiration is weak. Try repositioning the electrodes to improve the signal.
7.9.6 Changing Resp Wave Settings
WARNING
z
When monitoring in manual detection mode, make sure to check the respiration
detection level after you have increased or decreased the size of the respiration
wave.
In the [
Resp Waveform
] menu, you can:
Select [
Gain
] and then select an appropriate setting. The bigger the gain is, the larger the
wave amplitude is.
Select [
Sweep
] and then select an appropriate setting. The faster the wave sweeps, the
wider the wave is.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...