Alarm Messages
D-7
Measure
ment
Alarm message
A
B
Cause and solution
AG Change Watertrap
√
×
Wait until the change is completed.
AG Watertrap Type Wrong
√
×
Make sure that a correct watertrap has been
used.
O2 Accuracy Unspecified
×
√
N2O Accuracy
Unspecified
√
×
CO2 Accuracy
Unspecified
√
×
Enf Accuracy Unspecified
√
×
Iso Accuracy Unspecified
√
×
Sev Accuracy Unspecified
√
×
Hal Accuracy Unspecified
√
×
Des Accuracy Unspecified
√
×
RR Accuracy Unspecified
√
×
The measured value has exceeded the specified
accuracy range.
O2 Sensor Err
√
×
Galvanic O2 Sensor Err
√
×
AG Hardware Err
√
×
An error occurred to the module. Re-plug the
module or restart the monitor.
AG Airway Occluded
√
×
Check the airway and remove the occlusion.
AG Zero Failed
√
×
Re-plug the module or restart the monitor, and
then perform a zero calibration again.
AG Oxima Depletion Err
√
×
AG Oxima Depletion
Warn
√
×
Contact your service personnel and replace the
O2 battery.
AG Accuracy Err
√
×
The measured value has exceeded the specified
accuracy range.
RM No Sensor
√
×
RM Sensor Reversed
×
×
Check and reconnect the sensor.
Zero Failed
×
×
Perform a zero calibration again.
RM
RM Power Err
×
√
There is a problem with the power supply.
Re-plug the module or restart the monitor.
BIS High Imped.
√
×
BIS Sensor Off
√
×
Check and reconnect the BIS sensor.
BIS DSC Err
×
×
An error occurred to the DSC during receiving
signals. Check the DSC.
BIS
BIS DSC Malf
×
×
The DSC automatically shuts down as a result of
malfunction. Check the DSC.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...