Monitoring Cardiac Output
12-6
12.5 Measuring the Blood Temperature
As shown below, the blood temperature is measured with a temperature sensor at the distal
end of the catheter in the pulmonary artery. During C.O. measurements, blood temperature
alarms are suppressed to avoid false alarms. They will automatically recover as soon as the
C.O. measurements are completed.
12.6 Changing C.O. Settings
12.6.1 Setting the Temperature Unit
In the [
C.O. Setup
] menu, select [
Temp Unit
] to toggle between [
ºC
] and [
ºF
].
12.6.2 Setting the Interval between Measurements
To avoid inaccurate measurements, a certain period of time should be allowed for the blood
temperature to become stable before starting a new measurement. To set the interval
between two measurements, in the [
C.O. Setup
] menu, select [
Interval (s)
] and enter an
appropriate value. Every time when a measurement is completed, the monitor will count
down the remaining time and give corresponding prompt messages.
Balloon
Sensor
Pulmonary artery
Right artrium
Right ventricle
PA catheter
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...