Monitoring NIBP
9-8
An air tubing
A correct sized cylinder
Follow this procedure to perform the leakage test:
1. Set the patient category to [
Adult
].
2. Connect the cuff to the NIBP connector on the monitor.
3. Wrap the cuff around the cylinder as shown below.
4. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
NIBP Leakage Test
]. The NIBP display
shows [
Leakage Testing…
].
5. After about 20 seconds, the monitor will automatically deflate. This means the test is
completed.
6. If the message [
NIBP Pneumatic Leak
] is displayed, it indicates that the NIBP airway
may have leakages. Check the tubing and connections for leakages. If you ensure that
the tubing and connections are all correct, perform a leakage test again.
If the problem persists, contact your service personnel.
NOTE
z
The leakage test is intended for use to simply determine whether there are leakages
in the NIBP airway. It is not the same as that specified in the EN 1060-1 standard.
9.11 NIBP Accuracy Test
The NIBP accuracy test is required once every two years or when you doubt the measured
NIBP.
Tools required:
T-piece connector
Approprating tubing
Balloon pump
Monitor
Connector for
NIBP cuff
Air tubing
Cylinder
Cuff
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...