Maintenance
25-2
25.2 Service Tasks
The following tasks are for our qualified service professionals only. Contact a qualified
service provider if your patient monitor needs the following services. Clean and disinfect
equipment to decontaminate it before testing or maintaining it.
Maintenance schedule
Frequency
Safety checks according to
IEC60601-1
At least once every two years, after any repairs where the
power supply is replace, or if the patient monitor has been
dropped.
Performance assurance for all
measurements not listed below
At least once every two years, of if you doubt the measured
values.
Defibrillator synchronization
At least once every two years, or as needed.
NIBP leakage test
At least once every two years, or by your hospital policy.
NIBP accuracy test
At least once every two years, or by your hospital policy.
NIBP calibration
At least once every two years, or by your hospital policy.
CO
2
calibration and performance
test
At least once every two years, of if you doubt the measured
values.
AG calibration
At least once every two years, of if you doubt the measured
values.
AG preventative maintenance
(pump check, internal fan check,
ambient pressure check and so forth
as described in the service manual)
At least once every two years, of if you doubt the measured
values.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...