Other Functions
22-2
22.3 Nurse Call
The patient monitor also provides nurse call signals to a nurse call system connected to the
monitor via the Auxiliary Output Connector. To obtain nurse call signals, connect a nurse call
system to the monitor and then follow this procedure:
1. Select
[Main Menu]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password.
2. Select
[
Others >>
] to access the [
Others
] menu.
3. Select
[
Auxiliary Output
] and then select [
Nurse Call
].
4. Select
[
Nurse Call Setup
] to change the nurse call settings as follows:
Select [
Signal Type
] and toggle between [
Pulse
] and [
Continuous
].
[
Pulse
]: the nurse call signals are pulse signals and each pulse lasts 1 second. When
multiple alarms occur simultaneously, only one pulse signal is outputted. If an
alarm occurs but the previous one is not cleared yet, a new pulse signal will also be
outputted.
[
Continuous
]: the nurse call signal lasts until the alarm ends, i.e. the duration of a
nurse call signal equals to that of the alarm condition.
Select [
Contact Type
] and toggle between [
Normally Open
] and [
Normally Closed
].
[
Normally Open
]: select if your hospital’s nurse call relay contact is normally
open.
[
Normally Closed
]: select if your hospital’s nurse call relay contact is normally
closed.
Select [
Alm Lev
] and set the alarm level for nurse call-triggering alarms.
Select [
Alarm Cat.
] and then select the category to which the nurse call-triggering
alarms belong.
Alarm conditions are indicated to nurses only when:
The nurse call system is enabled,
An alarm that meets your preset requirements occurs, and
The monitor is not in the alarm paused or silence status.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...