Monitoring ECG/Resp
7-7
7.3.4 Checking Paced Status
It is important to set the paced status correctly when you start monitoring ECG. The paced
symbol
is displayed when the [
Paced
]
status
is set to
[
Yes
]. The pace pulse markers “
︱
”
are shown on the ECG wave when the patient has a paced signal.
To change the paced status, you can select either:
the patient information area, or
[
Main Menu
]
→
[
Patient Setup
]
→
[
Patient Demographics
], or,
the ECG parameter window,
and then, select [
Paced
] from the popup menu and toggle between [
Yes
] and [
No
].
Warning
z
For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the
patient monitor could mistake a pace pulse for a QRS and fail to alarm when the
ECG signal is too weak.
z
For non-paced patients, you must set [Paced] to[No]. If it is incorrectly set to [Yes],
the patient monitor may be unable to detect premature ventricular beats (including
PVCs) and perform ST segment analysis.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...