Monitoring RM
17-3
17.2 Safety Information
WARNING
z
Check for leaks in the breathing circuit system, as they may significantly affect
respiratory mechanics readings.
z
Verify the Ventilation Mode. Improper settings of this function will cause incorrect
calculations of some parameters.
z
Match the airway adapter you select to the appropriate patient category. Improper
sensor selection may produce excessive ventilation resistance or introduce excessive
airway deadspace, as well as inaccurate scales and alarm limits.
z
Periodically check the flow sensor and tubing for excessive moisture or secretion
build-up and purge if necessary.
NOTE
z
To avoid the affects of excessive moisture in the measurement circuit, insert the
flow sensor airway adapter in the breathing circuit with the tubes upright.
z
Do not place the airway adapter between the endotracheal tube and an elbow as
this may allow patient secretions to block the adapter windows.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...