Alarms
6-7
6.5.4 Recording Alarms Automatically
When a measurement alarm occurs, automatic recording of all the measurement numerics and
related waveforms is possible when the measurement’s [
Alarm
] and [
Alm Rec
] are set on.
To switch automatic start of alarm recording on or off for individual measurements, you can
either:
1. In the setup menu for your desired measurement, select [
Alm Rec
] and toggle between
[
On
] and
[
Off
],
or
2. Select
[
Main Menu
]
→
[
Alarm Setup >>
]
→
[
Alarm Recordings >>
]. In the
[
Alarm
Recordings
]
menu, select your desired measurement and toggle between [
On
] and
[
Off
].
Additionally, you can change the length of the recorded waveforms. In the
[
Alarm Setup
]
menu, select [
Recording Length
] and toggle between [
8 s
], [
16 s
] and [
32 s
]:
[
8 s
]: 4 seconds respectively before and after the alarm trigger moment.
[
16 s
]: 8 seconds respectively before and after the alarm trigger moment.
[
32 s
]: 16 seconds respectively before and after the alarm trigger moment.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...