Product Specifications
A-10
Accuracy
3-, 5-, and 12-lead ECG: ±1 bpm or ±1%, whichever is greater.
12-lead analysis: ±2 bpm
Sensitivity 200
μ
V (lead II)
Response time to heart rate
change
Meets the requirements of ANSI/AAMI EC13-2002: Section
4.1.2.1 f).
From 80 to 120 bpm: less than 11 s
From 80 to 40 bpm: less than 11 s
Meets the requirements of ANSI/AAMI EC13-2002: section
4.1.2.1 g).
Time to alarm for tachycardia
Waveform 4ah - range:
4a - range:
4ad - range:
Waveform 4bh - range:
4b - range:
4bd - range:
11 s
11 s
11 s
11 s
11 s
11 s
Pace Pulse
Pace pulses meeting the following conditions are labelled with a
PACE marker:
Pace pulse markers
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
When tested in accordance with the ANSI/AAMI EC13-2002:
Sections 4.1.4.1 and 4.1.4.3, the heart rate meter rejects all
pulses meeting the following conditions.
Pace pulse rejection
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
ST Segment Analysis
Measurement range
-2.0 to 2.0 mV
Accuracy
-0.8 to 0.8 mV:
Beyond this range:
±0.02 mV or ±10%, whichever is greater.
Not specified.
Refreshing rate
10s
Arrhythmia Analysis
Events
Asystole, Vfib/Vtac, PVC, Couplet, VT>2, Bigeminy,
Trigeminy, R on T, Missed Beats, Tachy, Brady, PNC and PNP
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...