Alarm Messages
D-5
Measure
ment
Alarm message
A
B
Cause and solution
NIBP Overrange
√
×
The measured NIBP value is not within the
specified range.
NIBP Excessive Motion
√
×
Check the patient’s condition and reduce the
patient motion.
NIBP Cuff Overpress.
√
×
The NIBP airway may be occluded. Check the
airway and measure again.
NIBP Equip Err
√
×
NIBP Timeout
√
×
NIBP Measure Failed
√
×
An error occurred during NIBP measurement
and therefore the monitor cannot perform
analysis correctly. Check the patient’s condition
and NIBP connections, or replace the cuff.
NIBP Illegally Reset
√
×
An illegal reset occurred during NIBP
measurement. Check if the airway is occluded.
YY Sensor Off
√
×
IBP
YY represents an IBP label.
C.O.
BT Sensor Off
√
×
CO
2
Sensor Off
×
√
Check the sensor connection and reconnect the
sensor.
CO
2
Internal Comm Err
√
There is a problem with the
communication between the CO
2
module and the monitor.
CO
2
Sensor High Temp
×
×
Check, stop using or replace the sensor.
CO
2
Sensor Low Temp
×
×
Check, stop using or replace the sensor.
CO
2
Sensor Fault
×
×
Check, stop using or replace the sensor.
CO
2
Airway High Press.
×
×
CO
2
Airway Low Press.
×
×
An error occurred in the airway pressure. Check
the patient connection and patient circuit, and
then restart the monitor.
CO
2
High Barometric
Press.
×
×
CO
2
Low Barometric
Press.
×
×
Check the CO2 connections, make sure that the
monitor application site meets the requirements,
and check for special sources that affect the
ambient pressure. Restart the monitor.
CO
2
FilterLine Occluded
×
×
CO
2
Watertrap Occluded
×
×
The airway or watertrap was occluded. Check
the airway and remove the occlusion.
CO
2
No Watertrap
×
√
Check the watertrap connections.
CO
2
Signal Saturated
√
×
CO
2
Signal Low
√
×
CO
2
Signal Too Low
√
×
CO2
CO
2
Signal Noisy
√
×
There is a problem with the CO2 signal quality.
Check the CO2 connections.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...