Monitoring ECG/Resp
7-11
7.5.6 Switching Defibrillator Synchronization On/Off
To switching defibrillator synchronization on or off:
1. Select the ECG parameter window to enter its setup menu.
2. Select
[
Defib. Sync
] and toggle between [
On
] and [
Off
].
If a defibrillator is connected, a defibrillator synchronization pulse (100 ms, +5V) is
outputted through the Defib. Sync Connector every time when the patient monitor detects an
R-wave. When [
Dfib. Sync
] is set on, the message [
Defib Sync On
] is displayed in the
technical alarm area.
WARNING
z
Improper use of a defibrillator may cause injury to the patient. The user should
determine whether to perform defibrillation or not according to the patient’s
condition.
z
Before defibrillation, the user must ensure both defibrillator and monitor has
passed the system test and can be safely used jointly.
z
Before defibrillation, make sure that [Defib. Sync] is set to [On] and the [Filter] is
set to [Diagnostic].
z
After defibrillation is finished, set [Defib. Sync] to [Off] and select the filter mode
as required.
7.5.7 Changing ECG Wave Settings
To change a wave’s settings, select the wave you want to change to enter its lead menu. In the
lead menu:
If the wave is too small or clipped, you can change its size by selecting an appropriate
[
Gain
] setting. If you select [
Auto
] from [
Gain
], the patient monitor will automatically
adjusting the size of the ECG waves. In [
Norma
l] screen, only the selected ECG wave’s
size is adjusted. In other screens, all ECG waves’ size is adjusted simultaneously.
You can change the wave sweep speed by selecting [
Sweep
] and then selecting the
appropriate setting.
You can adjust a wave’s position by adjusting the
or
button beside [
Up and
Down
] button.
You can restore a wave to its original position by selecting [
Back to Default
].
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...