Monitoring NIBP
9-9
Metal Vessel (volume 500±25 ml)
Reference manometer (calibrated with accuracy higher than 1 mmHg)
Follow this procedure to perform the accuracy test:
1. Connect the equipment as shown.
2. Before inflation, the reading of the manometer should be 0. If not, disconnect the airway
and reconnect it unit the readings is 0.
3. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
NIBP Accuracy Test
].
4. Compare the manometer values with the displayed values. The difference between the
manometer and displayed values should be no greater than 3 mmHg.
5. Raise the pressure in the metal vessel to 50 mmHg with the balloon pump. Repeat step 3
and 4.
6. Raise the pressure in the metal vessel to 200 mmHg with the balloon pump. Repeat step
3 and 4.
If the difference between the manometer and displayed values is greater than 3 mmHg,
contact your service personnel.
9.12 Calibrating NIBP
NIBP is not user-calibrated. Cuff-pressure transducers must be verified and calibrated at least
once a year by a qualified service professional. Contact your service personnel when a
calibration is necessary.
Monitor
Connector for
NIBP cuff
Manometer
Tubing
Balloon pump
Metal vessel
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...