Monitoring Carbon Dioxide
13-3
13.3 Preparing to Measure CO
2
13.3.1 Using a Sidestream CO
2
Module
1. Attach the watertrap to the module and then connect the CO
2
components as shown
below.
2. By default, the sidestream CO
2
module is in standby mode. The [
CO
2
Standby
]
message appears on the screen when the CO
2
module is plugged. To set the CO
2
module
to measure mode, Select the
hardkey on the module to switch to the [
Measure
]
mode, or select the CO
2
parameter window
→
[
Operating Mode
]
→
[
Measure
]. The
message [
CO2 Startup
] is displayed.
3. After start-up is finished, the CO
2
module needs time to warm up to reach the operating
temperature. The message [
CO
2
Sensor Warmup
] is displayed. If you perform CO
2
measurements during warm-up, the measurement accuracy may be compromised.
4. After warm-up is finished, you can perform CO
2
measurements.
NOTE
z
To extend the lifetime of the watertrap and module, disconnect the watertrap and
set the operating mode to standby mode when CO
2
monitoring is not required.
CAUTION
z
The watertrap collects water drops condensed in the sampling line and therefore
Watertrap
Watertrap fixer
Sampling line
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...