8-1
8
Monitoring SpO
2
8.1 Introduction
SpO
2
monitoring is a non-invasive technique used to measure the amount of oxygenated
haemoglobin and pulse rate by measuring the absorption of selected wavelengths of light.
The light generated in the probe passes through the tissue and is converted into electrical
signals by the photodetector in the probe. The SpO
2
module processes the electrical signal
and displays a waveform and digital values for SpO
2
and pulse rate. It provides four
measurements:
1. Pleth waveform (Pleth): visual indication of patient’s pulse.
2. Oxygen saturation of arterial blood (SpO
2
): percentage of oxygenated hemoglobin in
relation to the sum of oxyhemoglobin and deoxyhemoglobin.
3. Pulse rate (derived from pleth wave): detected pulsations per minute.
4. Perfusion indicator: the pulsatile portion of the measured signal caused by arterial
pulsation.
4
2
1 3
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...