Monitoring Carbon Dioxide
13-9
Select [
Sweep
] and the select the appropriate setting. The faster the wave sweeps, the
wider the wave is.
Change the size of the CO
2
waveform by adjusting the wave
[Scale
].
13.4.9 Setting the Automatic Standby Time
For the microstream CO
2
module, you can set a period of time after which the CO
2
module
enters the standby mode if no patient breath is detected. To set the standby time, in the [
CO2
Setup
] menu, select [
Auto Standby (min)
] and then select the appropriate setting.
13.5 Troubleshooting the Sidestream CO
2
Sampling
System
When the sampling system of the sidestream CO
2
module works incorrectly, check if the
sampling line is kinked. If not, remove it from the watertrap. If the monitor gives a message
indicating the airway still works incorrectly, it indicates that the watertrap must have been
blocked, and you should replace with a new one. Otherwise, you can determine that the
sampling line must have been blocked. Replace with a new sampling line.
13.6 Removing Exhaust Gases from the System
WARNING
z
Anesthetics: When using the Sidestream or Microstream CO
2
measurement on
patients who are receiving or have recently received anesthetics, connect the outlet
to a scavenging system, or to the anesthesia machine/ventilator, to avoid exposing
medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet
connector of the module.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...