Factory Defaults
C-8
C.11 CO
2
Setup
Factory default settings
CO
2
Setup
Adult
Pediatric
Adult
Alarm On
Alm Lev
Med
Alm Rec
Off
Humidity Compen
(side/microstream)
Wet
Press. Unit
mmHg
Max Hold (microstream)
20 s
Max Hold (mainstream)
10 s
Operating Mode (side/mainstream)
Standby
Operating Mode (microstream)
Measure
O2 Compen (side/mainstream)
Off
N2O Compen (sidestream)
Off
Des Compen (sidestream)
Off
Balance Gas (mainstream)
Room Air
Auto Standby (min) (microstream)
0
EtCO2 High
50
50
45
EtCO2 Low
15
20
30
FiCO2 High
4
4
4
awRR High
30
30
100
awRR High
8
8
30
Apnea Time
20 s
CO
2
Waveform
Wave Type
Draw
Sweep 25
mm/s