Monitoring BIS
16-8
16.8 Choosing the BIS Smoothing Rate
The smoothing rage defines how the monitor averages the BIS value. To choose the
smoothing rate:
1. Select the BIS parameter window to enter the [
BIS Setup
] menu.
2. Select
[
Smoothing Rate
] and then choose either:
[
15 s
]: this gives increased responsiveness to changes in the patient’s state.
[
30 s
]: this gives a smoother BIS trend with decreased variability and sensitivity to
artifacts.
16.9 Setting up the EEG Wave
Select the EEG wave to enter the [
BIS EEG Waveform
] menu. In this menu, you can:
Select [
Scale
] and then select the appropriate setting.
Select [
Sweep
] and then select the appropriate setting. The faster the wave sweeps, the
wider the wave is.