Monitoring IBP
11-7
Follow this procedure to perform the pressure calibration:
1. Disconnect the pressure transducer from the patient. Connect the 3-way stopcock, the
sphygmomanometer and the balloon pump through a T-shape connector, as shown
below.
2. Zero the transducer. After a successful zero, open the stopcock to the manometer.
3. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
Cal
.
IBP Press. >>
]. In the [
Cal. IBP
Press.
] menu, enter calibration values for calibrated pressures.
4. Inflate using the balloon pump until the mercury in the manometer reaches to the preset
calibration pressure value.
5. Adjust the preset calibration value until it equals to the reading on the manometer.
6. Select
the
[
Calibrate
] button to the right of the calibrated pressure. The monitor starts a
calibration.
7. When the calibration is completed, the message [
Calibration Completed!
] is displayed.
If the calibration failed, a prompt message will be displayed.
8. After the calibration is completed, disconnect the blood pressure tubing and the T-shape
connector.
Manometer
T-shape connector
3-way stopcock
Pressure transducer
Pressure adapter cable
IBP module
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...