Product Specifications
A-11
A.5.2 Resp
Technique Trans-thoracic
impedance
Lead
Options are lead I and II. The default is lead II.
Excitation current
<300 µA
Respiration impedance range
0.3 to 5
Ω
Baseline impedance range
200 to 2500
Ω
(using an ECG cable with 1k
Ω
resistance)
Differential input impedance
>2.5 M
Ω
Bandwidth
0.2 to 2 Hz (-3 dB)
Sweep speed
6.25, 12.5 or 25 mm/s
Respiration Rate
Measurement range
Adult:
Pediatric, neonate:
0 to 120 rpm
0 to 150 rpm
Resolution 1
rpm
Accuracy
7 to 150 rpm:
0 to 6 rpm:
±2 rpm or ±2%, whichever is greater
Not specified.
Apnea alarm time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s