Monitoring Cardiac Output
12-3
12.4 Setting Up the C.O. Measurement
WARNING
z
Use only accessories specified in this manual. Make sure that the accessories never
come into contact with conductive parts.
1. Connect the C.O. cable to the C.O. connector on the monitor.
2. Interconnect the C.O. module, catheter and syringe as shown below. Make sure that:
The module is securely inserted.
The PA catheter is in place in the patient.
The C.O. cable is properly connected to the module.
3. Check if the height and weight are appropriate for your patient in the [
Patient
Demographics
] menu. Change if necessary.
C.O. module
C.O. cable
Injectate
Syringe
In-line probe
PA catheter
Ballon inflation port
Ice water
Stopcock
Hole