Appendix 1 Error Messages
A
5
Appe
ndix
81
Perform initialization to start synchro-
nized measurements.
A setting was changed after starting syn-
chronous operation. Resynchronize by
pressing the [Synchronous Start] button
on the System screen of the master instru-
ment.
"14.2.4 Synchronized
Sampling Output
(SYNC.OUT)" (p. 384)
82
Maximum sampling rate is limited to 1
MS/s.
The sampling rate is limited when using
synchronized sampling.
83
No channel selected for use.
Select the channel(s) to use.
"4.2.1 Selecting Channels
to Use" (p. 86)
84
Measurement aborted due to save pro-
cessing delay.
With the Real-Time Saving function, mea-
surement is aborted if the recording speed
at the save destination is too slow. Select
a slower timebase or reduce the number
of channels used.
85
Invalid search condition.
Check the search criteria settings.
"8.14 Searching a Wave-
form" (p. 222)
86
Recording length is too long.
Check the recording length.
"4.2.4 Setting the Record-
ing Length (number of divi-
sions)" (p. 97)
87
Cannot load measurement waveform.
Load an index file (.RSI) created by the
Real-Time Saving function. If the RSI file
cannot be loaded, a measurement wave-
form file (.RSM) may be damaged or miss-
ing.
88
Function not available. (Roll,
Timebase2, Memory Division)
The Roll Mode, Timebase 2 and Memory
Division are not available when using the
Averaging function.
89
Function not available. (Trigger Priority,
Stop Trigger)
Trigger Priority and Stop Trigger are not
available when using the Averaging func-
tion.
90
Scanner module and Logic will not mea-
sure.
Scanner and Logic modules are disabled
available when using the Averaging func-
tion.
91
Wave Calculation will not automatically
execute on measuring.
Waveform calculations cannot be per-
formed while measuring using the Averag-
ing function. Calculations can be initiated
manually when measurement is finished.
92
AAF is disabled.
Anti-aliasing is not available when using
the Recorder function with external sam-
pling.
93
Measurement is not possible with the
current module configuration. Install no
more than four 8961s.
Up to four 8961 High Voltage Units (mod-
ules) may be used. Install no more than
four 8961s. (8861-50)
175
Internal temperature is rising. Please
confirm the use temperature and the in-
stallation environment.
Internal temperature is rising. Damage
may result if this condition continues.
Please confirm the use temperature and
installation environments.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......