11.4 Loading Data
295
11
Cha
pter 1
1 Sav
ing/Load
ing Dat
a & Managi
ng Files
Loadable settings data file:
See
"11.2 Data Capable of Being Saved & Loaded" (p. 267)
By loading an index file, data files are loaded as a batch.
To load waveform data in a batch
When waveform data is saved by Memory Division or in recording length divi-
sions, an index file is created along with the waveform data files.
By loading this index file, the waveform data files are loaded as a batch.
11.4.3 Loading Waveform Data
Operating Key
Procedure
1
Select the data to load (p. 292).
2
F1
Select
[Load].
A confirmation dialog box appears.
3
F1
Select
[Execute].
The selected waveform data is loaded on the in-
strument.
To cancel loading
Select
F2
[Cancel]
.
Loading Waveform Data
To open the screen: Press the
FILE
key
File screen
If
F1
[Load]
is not displayed, press the
F8
key to switch to
[Page 1/3]
.
When the waveform data is loaded, the waveform displayed currently on the
instrument is deleted. The loaded waveform and settings appear.
Index File
Extension
Description
IDX
The divided data files are loaded all at once.
This index is created when saving data after setting the recording length for [Di-
vision] on the Save Settings screen (unless [All Blocks] is selected for Memory
Division, in which case no IDX file is created).
See
"11.3.7 Automatically Saving Waveforms" (p. 282)
"11.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" (p. 285)
SEQ
(When using Memory Division with the Memory function)
To create an index file:
Enable Memory Division (set it to [ON]), set the target blocks on the Save Set-
tings screen to [All Blocks], and save.
See
"11.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" (p. 285)
RSI
(Real-Time Saving Function only)
Loads data measured with the Real-Time Saving function
To create an index file:
It is automatically created when measuring with the Real-Time Saving function.
See
"Chapter 9 Measuring with Real-Time Saving" (p. 235)
R_M
(REC&MEM Function only)
Loads data measured with the REC&MEM function
When measured with Memory Division enabled, all waveform data blocks are
loaded at once.
Index file creation:
The file is created automatically when saved with the REC&MEM function en-
abled.
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......