1.

 

Description. 

The 

WSD15TIIDR

  is  a 

datalogger

  with  3  input  channels  to  acquire,  Biaxial  inclination,  and  Outdoor  temperature,  with  storage 

functionality of samples acquired. 

 

 

2.

 

Configuration. 

Printed on the external label, the correspondence between the measured axes and the number of recorded measure channels is 
reported, by means of the three icons below:  

 

 

 

Picture 1 - Product Image

 

Picture 2 - WSD15TIIDR - Channels/axes correspondence

 

Summary of Contents for WSD15TIIDR

Page 1: ...ti it 1 11 1 Description The WSD15TIIDR is a datalogger with 3 input channels to acquire Biaxial inclination and Outdoor temperature with storage functionality of samples acquired 2 Configuration Printed on the external label the correspondence between the measured axes and the number of recorded measure channels is reported by means of the three icons below Picture 1 Product Image X X Picture 2 W...

Page 2: ... ELECTRONIC LEVEL feature Working on slab s regulation sensor position can be corrected and the device can be placed as vertical as possible NOTE antenna must be facing downwards 2 1 ELECTRONIC LEVEL feature Activate the axis reset procedure for channel 2 placing the WineCapKey in the upper left corner and counting 6 amber flashes Led will flash based on the measured tilt RED flashes or GREEN flas...

Page 3: ...nd instance of the command stores the current position as initial offset In case of an unintentional reset it is always possible to place again the tool near the mechanical zero using ELECTRONIC LEVEL feature and repeat the reset process WineCapManager software shows and if necessary reset the offset stored in the datalogger using a remote radio command ELECTRONIC LEVEL and RESET READING features ...

Page 4: ...is operation mode stand alone with data downloads through USB the connection with the PC and the WineCapManager running on it is necessary in order to modify the device s operation mode The sampling interval must be set with the device in STAND ALONE refer to Stand alone USB datalogger installation mode and automatically the device s clock is aligned with the PC s clock in order to assure the temp...

Page 5: ... it web www capetti it 5 11 4 Wireless device user interface The user interface consists of a virtual button that can be activated using the WineCapKey and of a two coloured led To give a command user must approach the WineCapKey to the device s sensible area and keep it in that position the following picture Picture 7 WineCapKey positioning shows device s sensible points Picture 7 WineCapKey posi...

Page 6: ...topped The sampling process and the radio are is turned off losing the connection to the network To reactivate a TEST command is necessary The STANDBY command must be given twice to confirm it at the first sequence the led flashes alternating RED and GREEN lights waiting for the second confirm sequence within 15 seconds At the command execution the led flashes as the STANDBY status refer to WIRELE...

Page 7: ...different command is associated to confirm the command the WineCapKey must be removed from the sensible area immediately after the number of flashes corresponding at the desired command The TEST corresponds to the second pulse and activate the Field Measurer function The device will give back the radio signal quality through led flashes WIRELESS MODE STATUS Table Flash count Wireless mode Status R...

Page 8: ...sh count Stand alone mode STATUS 1 green flash 2 seconds long ACTIVE 2 red flashes 2 seconds long STANDBY Sequence of red flashes short 2 seconds long short FACTORY RESET INVALID datalogger clock PC connection required 8 Shutting off Reactivating the device If the device is shut off and left unused for a long time you can issue the STANDBY command refer to Picture 8 Commands table Its corresponds ...

Page 9: ...mensions 100x100x80mm Weight 1 6000g Case material Aluminium Mounting Wall mounting with 2 gross setting Connections Wireless USB Inclination Transducer type Uniaxial Biaxial Triaxial MEMS technology 1 15 Dual Range Automatic Autoscale Inclination Measure range 1 15 Inclination Measure accuracy 0 5 of sample Inclination Measure resolution 0 001 1 range 0 01 15 range Inclination Mechanical zero adj...

Page 10: ..._R11 V4 7 CAPETTI ELETTRONICA s r l Strada Stratta 57 10090 CASTIGLIONE TORINESE TORINO ITALY VAT 04837130014 Phone 011 981 98 11 Fax 011 981 98 210 e mail info capetti it web www capetti it 10 11 10 Mechanical dimensions Picture 11 Mechanical dimensions ...

Page 11: ...11 65 CE that restricts the use of substances in the manufacturing of electronic devices The WEEE logo on the label indicates that this product is compliant with the WEEE EC Directive This symbol valid only in the European Union countries indicates that the product it is applied to MUST NOT be discarded with ordinary household or industrial waste but must be sent to an authorized reception point T...