8.14 Searching a Waveform
230
Operating Key
Procedure
6
(If searching for local maxima or minima)
Set the criteria for the local maximum or minimum value
(Filter).
CURSOR
F1 to F8
Move the cursor to the
[Filter]
item.
Set the criteria for the local maximum or mini-
mum value. (1div = 100 points)
Search
7
CURSOR
F1
Move the cursor to the
[Exec]
button.
Starts searching.
[Searching] appears on the Status bar, and the
locations where search criteria are satisfied are
displayed.
Search markers (
indicate locations where
search criteria are met.
To search for more instances (local maximum or mini-
mum values only)
F2
Select
[Next]
.
To move a cursor to a found location
F4
or
F5
Select
[Move Cur A]
or
[Move Cur B]
.
To insert an event mark at a location where search
criteria are met.
CURSOR
F6
Move the cursor to the
[Mark]
button.
Select
[Mark].
8
Close the search function.
F8
Select
[Close]
.
Off
When a value is larger (or smaller) than the
value of the data points immediately before
and after it, that value is considered to be a
local maximum (or minimum) (default set-
ting).
0.1 to
10.0 div
When a value is larger (or smaller) than the
values of all of the data points within the
specified range before and after it, that val-
ue is considered to be a local maximum (or
minimum).
S
Local maxima when the filter setting
is Off
Local maximum when the filter setting
is 0.1 div (0.1div=10 points)
10 points
10 points
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......