6.9 Trigger by Timer or Time Intervals (Timer Trigger)
162
Set this to record at fixed times.
Triggering occurs at the specified interval from the specified Start time until the
Stop time. Before setting, verify that the clock is set to the correct time. If not, set
the clock on the Environment (Env) Settings screen (p. 364).
6.9
Trigger by Timer or Time Intervals (Timer
Trigger)
1
2
Timer Trigger
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Trigger Settings screen
See
Screen Layout (p. 35)
When using the REC&MEM function:
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Status Settings screen
Operating Key
Procedure
1
Enable or disable the timer trigger.
CURSOR
Move the cursor to the
[Timer Trigger]
item.
F1 to F8
Enable or disable the timer trigger.
2
Set Start and Stop times (when [On] is selected).
CURSOR
Move to cursor to the
[Month], [Day], [Hour]
and
[Minute]
items to set recording Start and Stop
times.
F1 to F8
Set the date and time.
To set the current date and time:
Move the cursor to the
[Now]
button, and select
F1
.
To set only Start or Stop time:
Move the cursor to the
[Start]
or
[Stop]
button of
the setting you want to disable, and select
F1
3
Set the Interval.
(To apply a trigger through the specified interval, from
Start to Stop)
CURSOR
Move to cursor to the
[Day], [Hour], [Minute]
and
[Second]
items of
[Interval]
.
F1 to F8
Set the recording interval.
After pressing the
START
key, recording starts at the specified Start
time.
To stop recording early:
Press the
STOP
key.
Off
Timer triggering is disabled.
On
Timer triggering is enabled.
Start Time
Stop Time
T
T
T
Interval
Records the specified recording length
START Key
Start Time
Stop Time
T
T
T
Interval
Records the specified recording length
T
T
When the specified interval is shorter
than the specified recording length:
When the recording length exceeds the
specified interval
The next trigger is not applied until the
data for the specified recording length has
been acquired.
When the recording length exceeds the
stop time
Recording time depends on the operating
function.
"About Stop Time and Recording Length"
(p. 163)
START Key
3
When the interval is set to zero
If the [Repeat] trigger mode is selected, or
REC&MEM function is used, measure-
ment is repeated from Start to Stop times.
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......